II. Precautions
- Doppler imaging requires good two dimensional non-doppler, black and white images
- Optimize the view and center structures of interest before applying doppler
-
Cardiac Cycle timing
- Measurements (2D or M-Mode) are performed in relation to Cardiac Cycle (systole and diastole)
- Often this can be estimated by visualizing wall motion and valve position
- However, interpretation of doppler signals must be in the context of Cardiac Cycle timing
- Valvular defects have specific characteristics and timing in relation to the Cardiac Cycle
- Freeze the image, and use the cine feature to scroll back and forth through the images
- Attach ekg lead for rhythm strip to Ultrasound machine if doppler is being performed
- Not usually done in most Emergency Department Bedside Ultrasounds
- Systole will correlate with start of QRS Complex to start of T Wave
- Early systole will coincide with QRS Complex
- Mid systole will coincide with ST Segment
- Late systole will coincide with start of T Wave
- Diastole will correlate with the end of the T Wave to the start of the next QRS Complex
- Early diastole will coincide with the end of the T Wave (or U-Wave if present)
- Mid diastole will continue until the P Wave
- Late diastole will coincide with the P Wave and PR Interval
- Measurements (2D or M-Mode) are performed in relation to Cardiac Cycle (systole and diastole)
- Confirm abnormal findings in multiple views and with additional modalities (e.g. PW, CW, Color)
- Confirm the timing of the abnormality in terms of the Cardiac Cycle
- Check that the machine settings are appropriate (e.g. Color Scale)
- Recheck the abnormality in several views (just as a single EKG lead finding needs confirmation in other leads)
- Pressure waves may demonstrate physical findings
III. Background: Pulsed and Continuous Wave Doppler signals
- Doppler signals are only interpretable when flow is parallel to probe!
- Mitral valve and Tricuspid valve are parallel in Apical 4 Chamber View
- Aortic Valve is parallel in Apical 5 Chamber View
- Deviations from parallel flow underestimate velocity
- Small mis-measurements in velocity result in large deviations in pressure (velocity is squared)
- Small deviations (<20 degrees) from parallel flow measurement result in <10% error
- Pressures are measured across valve openings
- Measured peak velocity and mean velocity are converted to pressure using Bernoulli Equation
- Heart maintains constant volume from one cardiac chamber or vessel to the next for forward flow
- Velocity (and pressure gradient) increases at valve narrowing to maintain constant volume
- Simplified Bernouli Equation: Pressure Gradient = 4 * V^2
- Velocity of 2 m/s translates to a pressure gradient of 4*2^2=16 mmHg
- Velocity of 3 m/s translates to a pressure gradient of 4*3^2=36 mmHg
- Velocity of 4 m/s translates to a pressure gradient of 4*4^2=64 mmHg
- Velocity of 5 m/s translates to a pressure gradient of 4*5^2=100 mmHg
IV. Technique: Pulsed Wave Doppler (PW)
- Cursor is moved to a specific heart location to listen to a specific signal over a brief, fixed window
- The cursor location represents the Sample Volume, and is the only source of signals
- Indications
- Aortic Stenosis
- Measure peak velocity with Sample Volume at LVOT in apical 5 chamber view)
- Diastolic function
- Sample Volume at mitral valve, early diastolic inflow in apical 4 chamber view
- Peak Mitral E-Wave Velocity (cm/sec)
- E-Wave deceleration time
- E-Wave/A-Wave Ratio
- Aortic Stenosis
- Aliasing (signal noise) occurs when frequency is high and Nyquist limit is exceeded
- Occurs when target object frequency is more than half of pulse repetition frequency (PRF)
- Aliasing occurs when RBCs move faster than 1/2 of PRF (RBC velocity too high)
- Methods to reduce aliasing
- Switch to continuous wave (CW) doppler OR
- Place Sample Volume indicator at a shallower, decreased depth (closer to the probe) OR
- Increase the pulse wave (PW) doppler scale (display zooms out)
- Show more of the m/s axis (zero-baseline shifts up)
- Allows for higher velocity flow appearing on the graph
V. Technique: Continuous Wave Doppler (CW)
- Returns signal from all Red Blood Cells moving across the Ultrasound beam (line)
- Indications
- Evaluate valve stenosis and regurgitation (identifying peak and mean velocity and pressure)
- Mitral Stenosis (peak and mean velocity and pressure)
- Aortic Regurgitation (pressure half-time in msec) during diastole
- Estimate Right Ventricular Systolic Pressure or RVSP (and Pulmonary Artery Systolic Pressure)
- Uses even trace tricuspid regurgitation (present in most patients)
- RVSP = RA Pressure (estimated based on IVC) + 4*Vtr^2
- Left ventricular outflow tract velocity time integral (LVOT VTI)
- Aortic outflow velocity wave is traced to estimate peak and mean systolic pressures
- Evaluate valve stenosis and regurgitation (identifying peak and mean velocity and pressure)
VI. Technique: Color Flow Imaging (Color Doppler Imaging or CDI)
- Displays Blood Flow velocities (doppler shifts) in 2 dimensions, showing multiple locations of flow
- Pulsed wave values from a grid of locations within a color box (color sample volume)
- Provides a real-time spatial map of velocities (as represented by colors) in relation to 2D structures
- Color demonstrates relative velocities to one another
- Interpret color flow and direction of signal in relation to the Cardiac Cycle (e.g. Mitral Regurgitation in systole)
- Unlike Pulsed Wave Doppler and Continuous Wave Doppler, color flow does not result in specific velocity values
- Results are qualitative, not quantitative, and severity of findings is best evaluated with PW and CW
- Color flow doppler may be used in most views (even when not parallel to flow)
- BLUE is flow AWAY from probe and RED is flow TOWARDS probe (Mnemonic: BART)
- With apical 4 chamber view, normal mitral valve flow in diastole is toward the probe (red)
- With apical 5 chamber view, normal aortic valve flow in systole is away from the probe (blue)
- BLUE is flow AWAY from probe and RED is flow TOWARDS probe (Mnemonic: BART)
- Aliasing (signal noise) also occurs (as with PW) when frequency is high and Nyquist limit is exceeded
- Appears a mix of multiple random colors in the color box
- Aliasing occurs with higher red cell velocities, and therefore may be associated with more severe pathology
- See Pulsed Wave Doppler (PW) above regarding mechanism of aliasing and associated fixes
- Color gradient scale must be adjusted for accurate interpretation (as with PW doppler scale adjustments)
- Typically keep color scale between 50-60 and adjust up or down for specific findings, then reset scale back
- Adjust the color scale up (zooming out) to evaluate higher velocities (e.g. ventricles)
- Adjust the color scale down (zooming in) to evaluate lower velocities (e.g. atria in ASD evaluation)
- However, setting scale too low will exaggerate low flow signals and may appear as pathology
- Incorrectly set scale if not caught, may result even in unnecessary Valve Replacements
- In the Phen-Fen lawsuits, decreased color scale was among fraudulent techniques used
- Frame Rates
- Frame rates (and processing speed, quality) are decreased when color box (sample volume) is enlarged
- Trade-off of greater sample volume at the cost of reduced frame rates
- Keep frame rates (displayed on screen) above 20 hz
- Decrease the color box size to focus on specific regions of interest (ROI width)
- Frame rates (and processing speed, quality) are decreased when color box (sample volume) is enlarged
- Gain
- Adjust the color gain to help identify abnormalities (e.g. regurgitation jets)
- However, avoid adding too much gain that will create the appearance of abnormalities (as with decreasing scale)
- Indications
- Best at displaying valvular regurgitation (e.g. Mitral Regurgitation, Aortic Regurgitation)
- Demonstrates pathology even when not parallel to flow (unlike PW and CW)
- Used as a quick screening tool for pathology (e.g. PLAX View of Aorta outflow and mitral valve)
- Use PW or CW to quantify in parallel views (e.g. apical 5 chamber and apical 4 chamber)
- Best at displaying valvular regurgitation (e.g. Mitral Regurgitation, Aortic Regurgitation)
VII. References
- Jordan (2019) Cardiac Ultrasound Protocol Manual, Gulfcoast Ultrasound, p 13-22
- Reynolds (2018) The Echocardiographer's Pocket Reference, Arizona Heart Association, p. 323-4
- Palma, Bourque and Jordan (2019) Introduction to Adult Echo Ultrasound Conference, GulfCoast Ultrasound, St. Petersburg