II. Definition
- Thoracic-implanted, self-contained (with battery and circuitry)
- Cardiac sensing and pacing device indicated for Symptomatic Bradycardia or Syncope
III. Indications: Permanent Pacemaker - General
- Sinus Node Dysfunction with Symptomatic Bradycardia or Syncope
- Advanced second or third degree AV Block (see below)
- Chronic bifascicular block
- Acute Myocardial Infarction requiring ventricular pacing beyond acute phase
- Hypersensitive carotid sinus syndrome and Neurocardiogenic Syncope
- Cardiac Transplantation requiring post-operative pacing
- Symptomatic recurring SVT
- Congenital long QT Interval
- Cardiac Resynchronization Therapy with biventricular pacing
- Symptomatic Heart Block with Cardiomyopathy or Congenital Heart Disease
IV. Indications: Permanent Pacemaker - AV Block
- Class I Indications (helpful)
- Sick Sinus Syndrome
- Symptomatic Bradycardia with frequent sinus pauses (>3 seconds) associated with symptoms
- Symptomatic chronotropic incompetence (inadequate Heart Rate response to Exercise or activity)
- Second Degree AV Block with Symptomatic Bradycardia
- Third Degree AV Block with one associated condition
- Symptomatic Bradycardia
- Documented Asystole (3 seconds or greater)
- Catheter ablation of the AV junction
- Neuromuscular disorder with AV Block
- Myotonic muscular disorder
- Kearns-Sayre Syndrome
- Erb's Dystrophy (limb-girdle)
- Peroneal muscular atrophy
- Sick Sinus Syndrome
- Class IIa Indications (probably helpful)
- Sick Sinus Syndrome
- Symptomatic Bradycardia with Heart Rate <40 bpm but symptom association with Bradycardia unclear
- Idiopathic Syncope and Sinoatrial Node dysfunction identified on electrophysiologic studies
- Asymptomatic third degree AV Block
- Asymptomatic Type II second degree AV Block
- Asymptomatic Type I AV Block at His level
- First degree AV Block and Pacemaker syndrome symptoms
- Sick Sinus Syndrome
- Class IIb Indications (Possibly helpful)
- Sick Sinus Syndrome
- Chronic awake rate <40 bpm in minimally symptomatic patients
- Marked First Degree AV Block (>0.3 seconds) with CHF
- Hypersensitive carotid sinus syndrome with recurrent Syncope
- ACC/AHA recommends if associated with ventricular Asystole for 3 or more seconds
- Cochrane review does not find sufficient evidence supporting Pacemakers for Carotid Sinus Syncope
- Romme (2011) Cochrane Database Syst Rev (10): CD004194 [PubMed]
- Sick Sinus Syndrome
- Class III Indications (Not helpful, possibly harmful)
- Asymptomatic Bradycardia due to medication
- Asymptomatic First Degree AV Block
- Asymptomatic Type I AV Block limited to supra-His
- Transient AV Block secondary to resolving condition
- Drug toxicity
- Lyme Disease
VI. Description: Pacemaker Codes
- Background
- Most Pacemakers in the U.S. are DDD (dual lead, dual sensed, dual response of trigger and inhibit)
- Some Pacemakers placed outside the U.S. are VVI (ventricle paced, ventricle sensed, ventricle inhibited)
- Position 1 (chamber paced)
- V - Ventricle
- A - Atrium
- D - Dual (A and V)
- O - None
- Position 2 (chamber sensed)
- V - Ventricle
- A - Atrium
- D - Dual (A and V)
- O - None
- Position 3 (response to sensing)
- Position 4 (programmable functions and rate modulation)
- P - Programmable rate and output
- M - Muti-programmability of rate, output, sensitivity
- C - Communicating via telemetry
- R - Rate modulation
- O - None
- Position 5 (anti-tachyarrhythmia)
- P - Pacing (anti-tachyarrhythmia)
- S - Shock
- D - Dual (P and S)
- O - None
VII. Types: Pacemaker modes
- Synchronous mode
- Synchronous with heart beat
- Asynchronous mode (fixed mode)
- Asynchronous in regards to heart beat (no response to sensing)
- Pacemakers revert to asynchronous mode when exposed to a magnet
VIII. Types: Pacemaker Selection for Sinus Node Dysfunction
- No signs or future risks for impaired AV conduction
- Rate response: Rate-responsive atrial pacer (AAIR)
- No rate response: Atrial Pacemaker (AAI)
- Impaired AV Conduction and no AV synchrony needed
- Rate response: Rate-responsive dual chamber (DDDR)
- No rate response: Ventricular Pacemaker (DDD)
- Impaired AV Conduction and AV synchrony needed
- Tachyarrhythmia (e.g. PSVT)
- Rate response: Rate-responsive dual chamber (DDDR)
- No rate response: Dual chamber Pacemaker (DDD)
- No Tachyarrhythmia
- Rate response: Rate-responsive dual and mode switch
- No rate response: Dual chamber with mode switching
- Tachyarrhythmia (e.g. PSVT)
IX. Precautions
- Magnetic field exposure
- Pacemakers typically switch to asynchronous pacing at a set rate on exposure to magnetic field
- Cell phones should be held on the opposite side of body, away from Pacemaker
- Had been theoretical risk only in the past
- More powerful magnets are as of 2022 installed in phones for wireless charging (e.g. IPhone 12 MagSafe)
- Case reports of magnet triggered Syncope and other pacer adverse events are now more common
- Nadeem (2021) J Am Heart Assoc
- Other wearable and personal use devices (e.g. Electronic Cigarettes) may interfere with Pacemaker function
- Magnets may be helpful in some emergency settings (e.g. applied to AICD that is delivering inappropriate shocks)
- MRI scans have been historically contraindicated
- Devices manufactured after 2000 are considered safe for non-thoracic MRI (after first 6 weeks)
- Avoid MRI with 6 weeks of placement due to device dislodgement
- Set Pacemaker to asynchronous mode (reset after scan), and turn AICD devices off
- Pacemaker may distort images
- Typically performed at tertiary centers with emergency backup and temporary Pacemaker reprogramming
- Russo (2017) N Engl J Med 376(8): 755-64 +PMID:28225684 [PubMed]
- Nazarian (2013) Circ Arrhythm Electrophysiol 6(2): 419-28 [PubMed]
- Battery life remaining in Pacemaker
- Pacers are interrogated to determine remaining battery life
- When battery life drops below ERI (Elective replacement indicator), Heart Rate will be be fixed at a manufacturer-set rate
- When battery life drops below EOL (End of life), Heart Rate will be fixed at a different fixed manufacturer-set rate
- External Defibrillation, cardioversion and external pacing
- Electrical shock may theoretically damage Pacemaker
- Emergency Defibrillation may be performed without regard to Pacemaker
- Avoid applying elective cardioversion pads directly over Pacemaker
- Try to apply pads at least 10-15 cm away from the device
X. Imaging: Chest XRay
- Indications
- May identify misplaced or damaged lead
- Identifies the Pacemaker type via a radiodense device stamp (requires a high penetration Chest XRay)
- Interpretation - Normal Pacemaker wire configuration
- Most newer Pacemakers are dual lead (but some may have only a single lead to either atrium or ventricle)
- Pacemaker tip positioning - PA View (or AP View)
- Pacemaker leads should coss the midline (from the left chest wall implantation site to the right heart)
- Atrial leads should be directed slightly upwards
- Ventricular leads should be directed slightly downwards
- Pacemaker tip positioning - Lateral View (or Lateral View)
- Pacemaker leads should be directed towards the Sternum (anteriorly)
- As with PA view, atrial leads are directed upward and ventricular leads downward
XI. Diagnostics: Electrocardiogram
- Pacer spike
- Long, very narrow signal preceding a complex
- ST and T Waves
- Should always be discordant or opposite to the major QRS Complex (similar to a Left Bundle Branch Block)
- Discordance (or QRS Complex in the opposite direction as the ST Segment and T Wave) is normal
- Concordance (QRS and ST/T Wave in same direction) may suggest acute Myocardial Infarction
- Sgarbossa Criteria may distinguish Pacemaker altered ST-T morphology versus an acute Myocardial Infarction
- Applies to Left Bundle Branch Block induced with a right ventricular Pacemaker lead
- No guidelines directing Sgarbossa use in Pacemaker patients
- However original Sgarbossa study included a subset of Pacemaker patients
- Should always be discordant or opposite to the major QRS Complex (similar to a Left Bundle Branch Block)
-
QRS Complex
- Typically Left Bundle Branch Block (Pacemaker leads are typically placed in the right ventricle)
XII. Diagnostics: Magnet placed over Pacemaker
- Indicated if Pacemaker malfunction is suspected
- Perform an Electrocardiogram before and after applying magnet
- Before magnet application (synchronous mode)
- Pacing is typically rate responsive via Pacemaker sensing functionality
- After magnet application (asynchronous mode, fixed mode)
- Turns off Pacemaker sensing functionality
- Pacemaker will now pace at an intrinsic rate (80-100 depending on manufacturer)
- QRS Complexes present after magnet application suggests a failure of Pacemaker sensing
- No QRS Complexes after magnet applications may suggest a failure of Pacemaker capture
XIII. Diagnostics: Pacemaker Interrogation
- All Pacemakers can perform Electrocardiograms (may be turned off to conserve battery life)
- Electrocardiogram functionality is enabled during interrogation
- Pacemaker programming head is placed directly over Pacemaker and push interrogate button
- Identify the patient's underlying rhythm
- Are the pacer spikes atrial or ventricular?
- Is the rhythm Pacemaker dependent?
- If uncertain, Pacemaker rate can be slowed to see if patient's rate also drops
- Can the Pacemaker sense the heart rhythm?
- Can the Pacemaker pace the heart?
- Are there other rhythm problems?
- Some Pacemakers can store abnormal rhythm events (e.g. VT runs) with a date-time stamp
- Is the Pacemaker programmed correctly?
- ERI (Elective replacement indicator)
- Battery life remaining before replacement is needed
XIV. Management: Troubleshooting a malfunctioning Pacemaker
- See Cardiac Pacemaker Infection
- Evaluation
- See Chest XRay (above) for Pacemaker lead positioning and device stamp
- See Magnet Application (above)
- See Electrocardiogram (above) for Pacemaker-related findings
- See Pacemaker Interrogation (above)
- Pacemaker device malfunction categories
- Failure to capture (despite pacer spike)
- Pacemaker spike present without ensuing QRS Complex
- May indicate lead Fracture or dislodgement, or malfunction central to the device
- Lead displacement is the most common cause and typically occurs in first month
- Other causes include cardiac ischemia or infarction, and Electrolyte abnormalities (esp. Hyperkalemia)
- Failure to output (without pacer spike, includes oversensing, exit block)
- Pacer spikes are absent
- May indicate lead Fracture or dislodgement, battery depletion
- May indicate device oversensing
- Pacemaker inhibited by a Hyperacute T Wave, muscle Fasciculation, cell phone within 10 cm
- Misinterpreted as an intrinsic beat
- Applying a magnet over the Pacemaker is diagnostic and therapeutic
- May indicate exit block
- Altered interface between Pacemaker lead and endocardium
- Causes include MI, fibrosis, Hyperkalemia, Antiarrhythmics
- Failure to sense (undersensing)
- Pacemaker fails to sense intrinsic cardiac activity and paces as if intrinsic activity is absent
- Pacer spike appear regardless of intrinsic rhythm
- May indicate lead dislodgement, altered position, acute right ventricular infarction
- Risk of R on T phenomenon
- Response depends on Pacemaker mode and programmed sensing parameters
- Placing a magnet over the Pacemaker will disable the aberrant functionality until it may be repaired
- Failure to capture (despite pacer spike)
- Pacemaker Associated Dysrhythmia
- Pacemaker mediated Tachycardia (endless loop Tachycardia)
- Dual-Chamber Pacemaker mediated Tachycardia is a reentry Tachycardia
- Antegrade conduction via Pacemaker, retrograde conduction through AV Node
- Atrial sensing of retrograde impulses from the ventricle (e.g. PVC) trigger another atrial impulse
- Starts a cycle of Pacemaker impulses, followed by AV re-entry triggering another impulse
- Occurs if the Pacemaker postventricular atrial refractory period is too short
- EKG demonstrates a Wide Complex Tachycardia
- Retrograde P Waves may be present
- Ventricular pacer spikes are present, differentiating the ekg from Ventricular Tachycardia
- Treatment
- Apply magnet to Pacemaker to disable atrial sensing (and returning to a default pacer rate)
- Adenosine and AV Nodal blockers (e.g. Diltiazem, Beta Blockers) may also slow rate (as for AVRT)
- Dual-Chamber Pacemaker mediated Tachycardia is a reentry Tachycardia
- Runaway Pacemaker syndrome
- Rare, but life threatening
- A damaged Pacemaker theoretically could sporadically increase paced rates at well above 100
- Typically due to battery failure or Pacemaker damage (primarily older Pacemakers)
- Pacer spike bursts may trigger Ventricular Fibrillation
- Apply magnet to return Pacemaker to default rate
- Plan emergent replacement of the device
- Sensor-Induced Tachycardia
- Non-cardiac electrical stimulation triggers atrial-sensing and increases ventricular paced rate
- EKG demonstrates an inappropriate pacer rate
- Management
- Apply magnet to deactivate atrial sensing
- Eliminate non-cardiac stimuli
- References
- Berberian, Brady and Mattu (2024) Crit Dec Emerg Med 38(6): 14-5
- Pacemaker mediated Tachycardia (endless loop Tachycardia)
- Other Specific Pacemaker problems
- Dislodged/displaced or Fractured Pacemaker lead
- Pacemaker lead displacement is most common
- Pacemaker leads are most prone to Fracture at the lead insertion site and under the clavicle or first rib
- Pacemaker syndrome (up to 20% of cases)
- Loss of atrial capture with only ventricle paced (atria contracts against a closed AV valve)
- Presents with retrograde pulsations into the neck
- Atrial distention results in diuresis and Hypotension
- Loss of Preload may result in secondary Heart Failure
- Symptoms include weakness, Fatigue, Dizziness, Orthostasis, Dyspnea and Chest Pain
- Inadequate pacer energy
- Twiddler's Syndrome
- Pacemaker patients may manipulate the Pacemaker site (subconsciously)
- Results in lead retraction and displacement
- Presentations
- Risk factors
- Elderly or obese (loose subcutaneous Pacemaker pocket)
- Dementia, Obsessive Compulsive Disorder or Developmental Delay
- Diagnosis
- Electrocardiogram (evaluate for appropriate sensing)
- See Chest XRay below regarding lead positioning (also observe for Fractured leads)
- References
- Jhun and Shoenberger in Herbert (2015) EM:Rap 15(7): 17
- Dislodged/displaced or Fractured Pacemaker lead
- Data to have available when communicating with Cardiology about a patient with possible Pacemaker-related problem
- Pacemaker information
- Most patients will have a card with Pacemaker information (device type, interrogation phone numbers)
- Pacemaker type can also be read from a high penetration Chest XRay
- Is the device a Pacemaker or Defibrillator?
- How many wires to heart are present?
- When and where was the device implanted?
- Pacemaker information
XV. Precautions: Magnetic Resonance Imaging (MRI)
- Most Pacemakers can not undergo MRI
- Risk of Arrhythmia or capture loss
- Certain tertiary centers perform MRI on non-MRI conditional devices with electrophysiologist present
- Abandoned leads can not typically undergo MRI
- Lead super-heating and local myocardial wall ablation
- MR Conditional Devices and leads may undergo MRI under certain conditions
- Most manufacturers as of 2018 produce MR conditional devices
- Obtain Chest XRay in advance of MRI
- Requires Pacemaker temporary reprogramming (off or asynchronous mode with higher rates and outputs)
- Expect scatter on imaging from the device (for targets near the device)
- References
- Abresch (2018) Cardiac Arrhythmia Conference, UMN, Minneapolis
XVI. Complications: Acute complications associated with placement
- Pneumothorax or Hemothorax (complicates 1-3% of Pacemaker placements)
- Myocardial perforation (with associated Cardiac Tamponade risk)
- Brachial nerve injury
- Pacemaker Infection (1-7% risk)
- Acute lead dislodgment (2-4% risk)
XVII. Complications: Chronic complications
-
Atrial Fibrillation
- Increased risk with ventricular pacing (e.g. RV pacing)
XVIII. References
- Jones and Orman in Majoewsky (2012) EM:Rap 12(5): 4-6
- Mallemat, Swaminathan and Egan in Herbert (2014) EM:Rap 14(10): 5-7
- Vanlandingham (2015) Crit Dec Emerg Med 29(10): 2-14
- Bernstein (1987) Pacing Clin Electrophysiol 10:794-9 [PubMed]
- Denay (2014) Am Fam Physician 89(4): 279-82 [PubMed]
- Epstein (2008) J Am Coll Cardiol 51(21):e1-62 [PubMed]
- Gregoratos (1998) J Am Coll Cardiol 31:1175-209 [PubMed]
- Gregoratos (2005) Am Fam Physician 71:1563-70 [PubMed]